prBAS ISO 17862:2025

Surface chemical analysis — Secondary ion mass spectrometry — Linearity of intensity scale in single ion counting time-of-flight mass analysers


Opće informacije
Status:Projekt
Broj stranica:22
Metoda usvajanja:Korice
Jezik:engleski
Izdanje:2.
Nadnevak realizacije:20.11.2024
Predviđeni nadnevak naredne faze:10.03.2025
Tehnički komitet:BAS/TC 49, Кemijski inžinjering, laboratorijska oprema i kozmetika
ICS:
71.040.40, Hemijska analiza

Apstrakt
This document specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight (TOF) secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity nonlinearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead-time. The correction can increase the intensity range for 95 % linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction formulae have been shown to be valid.

Životni ciklus
...

Izvorni dokument i stepen usklađenosti
ISO 17862:2022, identičan

Veza sa BAS standardima

Radni materijal

Samo članovi tehničkog komiteta imaju pristup radnom materijalu. Ukoliko ste član, molimo prijavite se sa vašim nalogom i dobićete pristup dokumentima. Пријавите се