prBAS EN ISO 11137-1:2027
Sterilization of health care products - Radiation - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
General information
Status:Project
Number of pages:52
Adoption method:Proglašavanje
Language:engleski
Edition:2.
Realization date:22.10.2026
Forseen date for next stage code:31.10.2026
Technical committee:BAS/TC 25, Healthcare
ICS:
11.080.01, Sterilization and disinfection in general
11.080.99, Other standards related to sterilization and disinfection
11.080.99, Other standards related to sterilization and disinfection
Abstract
1.1 This document specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices.
NOTE Although the scope is limited to medical devices, this document can be applicable to other products and equipment.
This document covers radiation processes employing irradiators using:
a) the radionuclide 60Co or 137Cs;
b) a beam from an electron generator; or
c) a beam from an X-ray generator.
1.2 This document is not applicable to processes for inactivating viruses or the causative agents of spongiform encephalopathies, such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease.
NOTE For information on such processes, see ISO 22442-1, ISO 22442-2, ISO 22442-3, ISO 13022 and ICH Q5A.
1.2.1 This document does not specify requirements for designating a medical device as sterile.
NOTE Regional and national requirements can designate medical devices as sterile. See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.2 This document does not specify a quality management system for the control of all stages of production of medical devices.
NOTE It is not a requirement of this document to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices can require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.3 This document does not require that biological indicators be used for validation or monitoring of radiation sterilization, nor does it require that a pharmacopoeial test for sterility be carried out for product release.
1.2.4 This document does not specify requirements for occupational safety associated with the design and operation of irradiation facilities.
NOTE Regulations on safety requirements for occupational safety related to radiation can exist in some countries.
1.2.5 This document does not specify requirements for the sterilization of used or reprocessed devices.
Lifecycle
...
Original document and degree of correspondence
EN ISO 11137-1:2026, identical
ISO 11137-1:2025, identical
Relation to BAS standards
Work material
Only members of the technical committee have access to work material. If you are a members of this technical committee you need to login to view the documents. Login