prBAS EN ISO 11607-3:2027
Packaging for terminally sterilized medical devices - Part 3: Requirements for process development for forming, sealing and assembly
General information
Status:Project
Number of pages:30
Adoption method:Proglašavanje
Language:engleski
Edition:1.
Realization date:22.10.2026
Forseen date for next stage code:31.10.2026
Technical committee:BAS/TC 25, Healthcare
ICS:
11.080.30, Sterilized packaging
Abstract
This document specifies requirements for process development for forming, sealing and assembly of packaging for medical devices to be terminally sterilized, when utilizing heat sealing technologies.
This document recommends minimum heat sealing equipment features to support subsequent validation, process control and monitoring.
This document applies to both preformed sterile barrier systems and sterile barrier systems.
This document utilizes the sterile barrier system specification to develop the process specification using the principles of risk management.
This document is intended to be used prior to process validation.
NOTE ISO 11607-2 provides requirements for process specification and process validation.
Lifecycle
...
Original document and degree of correspondence
EN ISO 11607-3:2026, identical
ISO 11607-3:2026, identical
Work material